Transcatheter edge-to-edge repair in mitral regurgitation: A comparison of device systems and recommendations for tailored device selection. A systematic review and meta-analysis
Basel F. Alqeeqa, Mohammed Al‐Tawilb, Mohammed Hamama, Mohammad Aboabdoa, M. Elrayesc, Juergen Leickd, Mohamed Zeinahe f, Assad Haneyag, Amer Harkyf h✉
- aFaculty of Medicine, Islamic University of Gaza, Gaza, Palestine
- bFaculty of Medicine, Al-Quds University, Jerusalem, Palestine
Papers open inside SciRah from the original publisher or repository; SciRah never hosts them.
Abstract
BACKGROUND Mitral valve transcatheter edge-to-edge repair (M-TEER) is a minimally invasive method for the treatment of mitral regurgitation (MR) in patients with prohibitive surgical risks. The traditionally used device, MitraClip, showed both safety and effectiveness in M-TEER. PASCAL is a newer device that has emerged as another feasible option to be used in this procedure. METHODS We searched for observational studies that compared PASCAL to MitraClip devices in M-TEER. The electronic databases searched for relevant studies were PubMed/MEDLINE, Scopus, and Embase. The primary outcomes were technical success and the grade of MR at follow-up. Secondary outcomes included all-cause mortality, bleeding, device success and reintervention. RESULTS Technical success (PASCAL: 96.5% vs MitraClip: 97.6%, p = 0.24) and MR ≤ 2 at 30-day follow-up (PASCAL: 89.4vs MitraClip 89.9%, p = 0.51) were comparable between both groups. Both devices showed similar outcomes including all-cause mortality (RR: 0.68 [0.34, 1.38]; P = 0.28), major bleeding (RR: 1.87 [0.68, 5.10]; P = 0.22) and reintervention (RR: 1.02 [0.33, 3.16]; P = 0.97). Device success was more frequent with PASCAL device (PASCAL: 86% vs MitraClip 68.5%; P = 0.44), however, the results did not reach statistical significance. CONCLUSION Clinical outcomes of PASCAL were comparable to those of MitraClip with no significant difference in safety and effectiveness. The choice between MitraClip and PASCAL devices should be guided by various factors, including mitral valve anatomy, etiology of regurgitation, and device-specific characteristics.
TL;DRClinical outcomes of PASCAL were comparable to those of MitraClip with no significant difference in safety and effectiveness, and should be guided by various factors, including mitral valve anatomy, etiology of regurgitation, and device-specific characteristics. (Semantic Scholar)